GSPR compliance for medical devices

GSPR Compliance for Medical Devices: EU MDR Requirements & Guide

GSPR compliance for medical devices is a fundamental requirement under the European Union Medical Device Regulation (EU MDR) 2017/745. Manufacturers must demonstrate that their devices meet the applicable General Safety and Performance Requirements (GSPR) specified in Annex I before placing them on the EU market.

GSPR compliance is not simply a checklist exercise. Manufacturers must determine which requirements apply to their device, identify appropriate methods of demonstrating compliance, and maintain objective evidence within their technical documentation.

A well-structured GSPR assessment also creates a clear link between device risks, applicable standards, testing, clinical evidence, and technical documentation.

Talk to Our Regulatory Experts

What Is GSPR Compliance for Medical Devices?

GSPR refers to the General Safety and Performance Requirements established in Annex I of EU MDR 2017/745.

These requirements address areas including:

  • Safety and performance
  • Risk reduction
  • Chemical and physical properties
  • Infection and microbial contamination
  • Materials and biological safety
  • Devices incorporating medicinal substances
  • Devices with measuring functions
  • Protection against radiation
  • Software and electronic systems
  • Mechanical and thermal risks
  • Information supplied by the manufacturer

The requirements applicable to a particular device depend on its design, intended purpose, materials, technology, use environment, and associated risks.

Manufacturers must demonstrate conformity with the applicable GSPRs and provide appropriate evidence.

What Are the EU MDR GSPR Requirements?

Annex I contains the GSPRs that manufacturers must evaluate.

They can broadly be grouped into three areas:

Chapter I – General Requirements

These requirements address fundamental principles such as:

  • Safety and performance
  • Risk management
  • Risk reduction
  • Benefit-risk determination
  • Usability
  • Device lifetime
  • Transport and storage
  • Environmental conditions

Chapter II – Requirements Regarding Design and Manufacture

These requirements address specific characteristics of device design and manufacturing, including:

  • Chemical, physical and biological properties
  • Infection and microbial contamination
  • Manufacturing processes
  • Devices incorporating medicinal substances
  • Devices composed of substances
  • Devices with measuring functions
  • Protection against radiation
  • Electronic programmable systems and software
  • Mechanical and thermal risks

Chapter III – Information Supplied With the Device

Manufacturers must also consider requirements for:

  • Labels
  • Instructions for use
  • Identification information
  • Warnings
  • Precautions
  • Symbols
  • Residual risks
  • Other information supplied with the device

The exact GSPR assessment should always be performed against the applicable provisions of Annex I, rather than relying on a generic checklist.

How Do You Demonstrate GSPR Compliance?

The basic process for GSPR compliance documentation can be divided into several steps.

Step 1: Review Every Applicable GSPR

Start by reviewing the requirements in Annex I and determine whether each requirement is:

  • Applicable
  • Not applicable
  • Partially applicable

If a requirement is considered not applicable, the manufacturer should provide a documented justification supported by appropriate technical or scientific reasoning.

For example, a requirement relating to a particular type of biological interaction may not apply to a device with no relevant patient contact. The justification should be specific to the device rather than simply stating “not applicable.”

Step 2: Identify the Method of Compliance

For every applicable GSPR, identify how conformity will be demonstrated.

Possible methods include:

  • Design verification
  • Design validation
  • Laboratory testing
  • Clinical evaluation
  • Risk management
  • Biocompatibility assessment
  • Usability engineering
  • Software verification and validation
  • Sterilization validation
  • Packaging validation
  • Electrical safety testing
  • Performance testing

The compliance method should be appropriate to the specific requirement and device.

Step 3: Identify Applicable Standards and Common Specifications

Manufacturers should identify relevant harmonized standards, Common Specifications where applicable, and other appropriate technical specifications used to demonstrate conformity.

Examples may include standards relating to:

  • Risk management
  • Quality management
  • Biocompatibility
  • Usability engineering
  • Sterilization
  • Packaging
  • Electrical safety
  • Software
  • Particular device categories

However, manufacturers should not simply list every standard they use. Each standard should have a clear connection to the device and the requirement it helps address.

Step 4: Link GSPR Requirements to Objective Evidence

Each applicable GSPR should be supported by appropriate evidence.

For example:

GSPR area

Compliance method

Supporting evidence

Risk reduction

Risk management

Risk management file

Biological safety

Biocompatibility assessment

Biological evaluation report

Sterility

Sterilization validation

Validation report

Electrical safety

Testing

Test report

Software safety

Software V&V

Software documentation

Clinical safety/performance

Clinical evaluation

CER

Labeling

Label review

Labeling/IFU

Usability

Usability engineering

Usability file

This evidence should be traceable to the manufacturer’s technical documentation.

What Is a GSPR Compliance Matrix?

A GSPR compliance matrix is one of the most useful tools for organizing GSPR compliance.

The matrix typically includes:

GSPR

Applicable?

Compliance method

Standard/specification

Evidence

Document reference

GSPR 1

Yes

Risk management

ISO 14971

Risk Management File

RMF-001

GSPR 2

Yes

Design verification

Applicable standard

Test Report

TR-001

GSPR 10

Yes

Biocompatibility

ISO 10993 series

Biological Evaluation

BER-001

GSPR 23

Yes

Label/IFU review

Applicable labeling requirements

IFU

IFU-001

The matrix creates a traceable relationship between the MDR requirement and the evidence demonstrating conformity.

It can also help identify gaps before technical documentation is submitted to a Notified Body.

Start Your GSPR Assessment

Identify applicable safety and performance requirements and determine the evidence needed for compliance.

GSPR Compliance and Risk Management

GSPR compliance should not be developed independently from risk management.

The EU MDR requires manufacturers to implement a risk management system, and risk-control measures can provide important evidence for demonstrating conformity with applicable GSPRs.

A practical relationship is:

GSPR → Hazard/Risk → Risk Control → Verification → Evidence

For example, if a device presents a mechanical hazard, the manufacturer should identify the associated risk, establish appropriate controls, verify those controls, and reference the resulting evidence within the GSPR documentation.

This integrated approach makes the technical documentation more consistent and easier to review.

GSPR Compliance and State of the Art

Manufacturers should consider the state of the art when selecting methods and standards for demonstrating conformity.

Using an outdated test method or inappropriate standard can weaken the compliance justification even when documentation appears complete.

Therefore, manufacturers should periodically review:

  • Applicable harmonized standards
  • Common Specifications
  • Scientific and technical developments
  • Relevant guidance
  • Industry practices
  • Device-specific safety expectations

The goal is to demonstrate that the selected compliance approach is appropriate for the device and reflects current knowledge and technology.

How Operon Strategist Supports GSPR Compliance

Developing GSPR compliance documentation requires regulatory knowledge and coordination across multiple technical disciplines.

Operon Strategist supports medical device manufacturers with:

  • EU MDR GSPR assessment
  • GSPR compliance matrix preparation
  • Technical documentation
  • Risk management
  • Applicable standards identification
  • Clinical evaluation support
  • Biocompatibility documentation
  • Usability engineering
  • Labeling and IFU review
  • CE Marking support
  • Notified Body submission support

Our experts help manufacturers establish a clear connection between each applicable GSPR and the evidence demonstrating conformity.

Assess Your GSPR Compliance

Get expert support to identify applicable EU MDR requirements, map evidence, and strengthen your technical documentation for CE marking.

FAQ's

GSPR compliance means demonstrating that a medical device meets the applicable General Safety and Performance Requirements in Annex I of EU MDR 2017/745 through appropriate technical, clinical, and quality evidence.

The GSPR requirements are defined in Annex I of Regulation (EU) 2017/745, the EU Medical Device Regulation.

The MDR requires manufacturers to demonstrate conformity with applicable GSPRs and maintain appropriate technical documentation. A document specifically titled a “GSPR compliance matrix” is a commonly used method for organizing this evidence and demonstrating traceability.

A typical matrix identifies the GSPR, applicability, method of demonstrating conformity, applicable standards or specifications, supporting evidence, and relevant technical-document references.

The manufacturer should document why the requirement is not applicable to the specific device and provide an appropriate technical or scientific justification.